User Contributed MET/CAL PROCEDURE ============================================================================== INSTRUMENT: WESTON 7060 INSTRUMENT: GUILDLINE 9574 DATE: 22-Apr-87 AUTHOR: User Contributed REVISION: V 4.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 51 NUMBER OF LINES: 96 CONFIGURATION: Fluke 5100B Fluke 8506A CONFIGURATION: ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP Allow minimum warm-up of 10minutes 1.001 DISP Set UUT to 10 VDC Range and short the input terminals 1.002 EVAL Does the UUT indicate Zero +/-.00001? 2.001 DISP Remove the short and connect test cables. 2.002 ASK+ R N M A L T W 2.003 HEAD DC VOLTAGE TESTS 2.004 DISP Set UUT to read at 6X9 scale length with filter in. 2.005 5100 0.1 0.1V 0.012% E 2W 3.001 5100 1 1V 0.0066% E 2W 4.001 5100 1 -1V 0.0066% E 2W 5.001 5100 10 10V 0.0066% E 2W 6.001 5100 10 -10V 0.0066% E 2W 7.001 5100 100 100V 0.0066% E 2W 8.001 5100 100 -100V 0.0066% E 2W 9.001 5205 1000 1000V .0066% E 2W 10.001 HEAD DC CURRENT TESTS 10.002 DISP Set scale length to 5X9 with filter out. 10.003 5100 100 10mA 0.05% 2W 11.001 5100 100 100mA 0.05% 2W 12.001 5100 1 1A 0.2% 2W 13.001 HEAD RESISTANCE TESTS 13.002 DISP Set T.I. to read 4-WIRE RESISTANCE with filter in 13.002 DISP and a 6x9 scale length 13.003 RESF 1 1KZ 0.003% 4W 14.001 RESF 10 10KZ 0.0026% 4W 15.001 DISP Switch to 2-wire Ohms Function. 15.002 RESF 100 100KZ 0.0026% 2W 16.001 RESF 1 1MZ 0.0029% 2W 17.001 RESF 10 10MZ 0.03% 2W 18.001 OPBR Does UUT have Average AC Volts Option? 18.002 JMPF 33.001 18.003 HEAD AVERAGE AC VOLTAGE TESTS 18.004 DISP Set UUT to AC Volts, 5X9 scale and filter in. 18.005 5100 0.1 0.1V 0.09% 50H 2W 19.001 5100 0.1 0.1V 0.09% 5KH 2W 20.001 5100 0.1 0.1V 0.23% 20KH 2W 21.001 5100 1 1V 0.09% 50H E 2W 22.001 5100 1 1V 0.09% 5KH E 2W 23.001 5100 1 1V 0.23% 20KH 2W 24.001 5100 10 10V 0.12% 50H E 2W 25.001 5100 10 10V 0.12% 5KH E 2W 26.001 5100 10 10V 0.33% 20KH 2W 27.001 5100 100 100V 0.12% 50H E 2W 28.001 5100 100 100V 0.12% 5KH E 2W 29.001 5100 100 100V 0.33% 20KH 2W 30.001 5100 1000 500V 0.12% 50H E 2W 31.001 5205 1000 500V 0.12% 5KH 2W 32.001 5205 1000 500V 0.33% 20KH 2W 33.001 OPBR Does the UUT have RMS VOLTAGE OPTION? 33.002 JMPF 49.001 33.003 HEAD RMS VOLTAGE TESTS 33.004 DISP Set UUT to AC Volts Function, 5X9 scale with filter in 33.005 5100 0.1 0.1V 0.25% 50H 2W 34.001 5100 0.1 0.1V 0.25% 5KH 2W 35.001 5100 0.1 0.1V 0.65% 20KH 2W 36.001 5100 1 1V 0.25% 50H 2W 37.001 5100 1 1V 0.25% 5KH 2W 38.001 5100 1 1V 0.65% 20KH 2W 39.001 5100 1 1V 0.65% 40KH 2W 40.001 5100 10 10V 0.3% 50H 2W 41.001 5100 10 10V 0.3% 5KH 2W 42.001 5100 10 10V 0.75% 20KH 2W 43.001 5100 100 100V 0.3% 50H 2W 44.001 5100 100 100V 0.3% 5KH 2W 45.001 5205 100 100V 0.75% 20KH 2W 46.001 5205 1000 500V 0.3% 50H 2W 47.001 5205 1000 500V 0.3% 5KH 2W 48.001 5205 1000 500V 0.75% 20KH 2W 49.001 HEAD AC CURRENT TESTS 49.002 DISP Set UUT to AC Current, 5X9 scale with filter in 49.003 5100 100 10mA 0.12% 50H 2W 50.001 5100 100 100mA 0.12% 400H 2W 51.001 5100 100 100mA 0.12% 5KH 2W END